Chapter 5 Examining Spectrum Data

5-10

Applied Biosystems

5

6.

Select the Peak Weighting Factor. If the calibration
includes more than two points, you can apply the following
weighting factors to fit points to the curve:

•

None—All peaks weighted equally

•

Inverse Width—Narrower peaks are weighted
more than broader peaks

•

Height—More intense peaks are weighted more
than less intense peaks

Manually matching

peaks

7.

To manually select the reference mass for a peak,
right-click-drag over the peak of interest.

The Select or Create Reference Peak Information
dialog box (Figure 5-3) is displayed and lists all masses
in the selected reference file. The selected entry is the
nearest match in the calibration reference file to the
selected peak.

Figure 5-3 Select or Create Reference Peak

Information Dialog Box