Chapter 2 Using Chromatogram and Spectrum Windows

2-12

Applied Biosystems

2

5.

Set the parameters described below as needed:

6.

Click OK.

Parameter

Description

Scaling Mode

Display Relative

Autoscales the trace to the largest peak in the selected range.

Base Peak
Relative

Autoscales the trace to the base peak in the entire range, not the
selected range. Displays a right-axis label with the base peak
intensity.

NOTE: To turn off the right axis, select Graphic Options from
the Display menu, click the Graph #1 Setup tab, then deselect
the Show Right Y-Axis check box.

Absolute Value

Sets the trace to the Y Display Range you enter in the Y Display
Range From/To boxes.

Display Min/Max

Sets the trace display to the minimum and maximum Y values.

Minimum Absolute Max Y

Use Limit

Sets the minimum value for Y-axis scaling. Useful to maintain
relative scaling of peaks and to prevent autoscaling noise to full
scale.

Y Display Range

From/To

Sets the range for scaling. If Display Relative or Base Peak
Relative is selected, range is in percent.

If Absolute Value is selected, range is in counts.