Chapter 8 Viewing Voyager PSD Data

8-18

Applied Biosystems

8

Solving and

plotting

After matching peaks, click Solve and Plot.

The calibration statistics are displayed in the Result tab of the
Output window and the calibration constants are applied to
the spectrum displayed

If you calibrate more than one time, subsequent calibration
statistics are added to the end of the list in the Output window.
Older calibration statistics are listed at the top of the list. Use
the scroll bar to view newer statistics at the bottom of the list.

The spectrum is calibrated and displayed with an MC trace
label. The calibration statistics are displayed in the Output
window.

Applying new

constants to the

data file

To apply the calibration constants to each spectrum (all
segments) in the data file, click Apply Calibration.

All spectra in the data file are calibrated and displayed with an
MC trace label. The calibration constants are saved with the
data file. Each spectrum in the data file is calibrated when
displayed.

The software uses different values in the calibration equation
calculation based on the type of calibration performed.