Chapter 3 Peak Detection and Labeling

3-72

Applied Biosystems

3

The following table lists the default settings in the .SET files
provided for spectra.

Parameter

Mariner

Spectrum

MARINER.S

ET

Voyager Spectrum

Linear

VOYAGER

LINEAR.SET

Reflector

VOYAGER

REFLECTOR

.SET

PSD

VOYAGER

PSD.SET

Basic Settings (spectrum data)

%Base Peak Intensity

0.25

0

0

0

%Max Peak Area

0.25

2

2

2

Use Resolution Dependent
Settings

ON

ON

ON

OFF

Mass Resolution

5,000

2,000

10,000

N/A

Peak Processing

Integration Baseline Setting

V-to-V

V-to-B

V-to-B

V-to-B

Centroid%

50

50

50

50

Max. Charge State

6

1

1

1

Max. Isotopes#

6

5

10

5

Min. Intensity%

5

10

10

10

Max. Intensity%

100

100

100

100