Chapter 7 Data Explorer Examples

7-8

Applied Biosystems

7

7.1.3 Determining if a Peak

is Background Noise

Overview

To determine if spectral peaks represent low-level
components or if they are due to solvent contribution, you can:

•

Subtract the spectral peaks from the chromatogram

•

Create an extracted ion chromatogram for the spectral
peaks

Subtracting

spectral peaks

The following example uses the chromatograms and spectra
from Section 7.1.2, Deconvoluting and Evaluating Unresolved
Chromatographic Peaks
. It illustrates how to subtract baseline
and determine if the spectral peak observed at 391 Da
(Figure 7-5 on page 7-7) is eliminated.

To subtract spectra:

1.

Activate the first extracted ion chromatogram in
Chromatogram window (see Figure 7-4 on page 7-6).

2.

From the Process menu, select Add/Subtract Spectra.

3.

Click

, then right-click-drag over the peak in the

extracted ion chromatogram.

The spectrum range is displayed in the Spectra To Be
Added list in the Add and Subtract Spectra dialog box
(Figure 7-6).

4.

In the Add and Subtract Spectra dialog box, click
anywhere in the Spectra To Be Subtracted list.

5.

In the TIC, right-click over a region of baseline that does
not contain any peaks.

NOTE: When selecting spectra to subtract, select from
the TIC. The baseline in the extracted ion
chromatograms is for a selected mass and does not
represent all components which may be present.

Advertising