Chapter 3 Peak Detection and Labeling

3-60

Applied Biosystems

3

Charge state

not displayed

If no charge is displayed, there are a few possible causes:

•

Peaks are more than 1 Da apart.

•

Filter width is set too high to detect other isotope peaks.

•

The maximum charge state for charge state
determination is set lower than the charge state of the
peak.

•

Charge state determination parameters are set such that
peaks are determined to have no charge. See
Section 3.2.5, Charge State Determination and
Examples
.

•

Peaks are not from the same isotope species.

NOTE: If you set %Base Peak Intensity or %Max Peak
Area too low (described on page 3-20)
, noise peaks
can be detected, and charge states can be incorrectly
labeled.