3 calibrating a psd spectrum, Calibrating a psd spectrum – HP Data Explorer 4 Series User Manual

Page 348

Advertising
background image

Chapter 8 Viewing Voyager PSD Data

8-10

Applied Biosystems

8

8.3 Calibrating a PSD Spectrum

NOTE: Multi-point calibration yields higher mass accuracy
than one-point calibration.

This section includes:

•

Checking peak detection

•

Calibrating

•

Creating PSD .CAL files and applying to other data files

•

Creating PSD calibration reference (.REF) files

•

Changing the precursor mass

When to use this

procedure

Use this procedure to:

•

Generate a PSD calibration (.CAL) file from a known
standard analyzed in PSD mode. You can then export the
PSD .CAL file and include the PSD .CAL file in a PSD
instrument settings (.BIC) file you use to acquire unknown
PSD samples.

•

Internally calibrate an unknown PSD data file by
specifying known monoamino acid fragment ions as the
peaks to match during calibration.

Overview of

creating a PSD

.CAL file

To create a PSD .CAL file:

•

Acquire a peptide standard with a known sequence in
PSD mode.

•

Check peak detection as described in Section 8.3.1,
Checking Peak Detection.

•

Create a calibration reference (.REF) file as described in
Section 8.3.4, Creating PSD
Calibration Reference (.REF) Files.

•

Internally calibrate the PSD .DAT file as described in
Section 8.3.2, Calibrating. Specify the calibration
reference (.REF) file created in Section 8.3.4, Creating
PSD Calibration Reference (.REF) Files,
when
calibrating.

•

Export the calibration (.CAL) file from the PSD .DAT file to
use when acquiring unknown samples in PSD mode.