3 calibrating a psd spectrum, Calibrating a psd spectrum – HP Data Explorer 4 Series User Manual

Page 348

Advertising
background image

Chapter 8 Viewing Voyager PSD Data

8-10

Applied Biosystems

8

8.3 Calibrating a PSD Spectrum

NOTE: Multi-point calibration yields higher mass accuracy
than one-point calibration.

This section includes:

Checking peak detection

Calibrating

Creating PSD .CAL files and applying to other data files

Creating PSD calibration reference (.REF) files

Changing the precursor mass

When to use this

procedure

Use this procedure to:

Generate a PSD calibration (.CAL) file from a known
standard analyzed in PSD mode. You can then export the
PSD .CAL file and include the PSD .CAL file in a PSD
instrument settings (.BIC) file you use to acquire unknown
PSD samples.

Internally calibrate an unknown PSD data file by
specifying known monoamino acid fragment ions as the
peaks to match during calibration.

Overview of

creating a PSD

.CAL file

To create a PSD .CAL file:

Acquire a peptide standard with a known sequence in
PSD mode.

Check peak detection as described in Section 8.3.1,
Checking Peak Detection.

Create a calibration reference (.REF) file as described in
Section 8.3.4, Creating PSD
Calibration Reference (.REF) Files.

Internally calibrate the PSD .DAT file as described in
Section 8.3.2, Calibrating. Specify the calibration
reference (.REF) file created in Section 8.3.4, Creating
PSD Calibration Reference (.REF) Files,
when
calibrating.

Export the calibration (.CAL) file from the PSD .DAT file to
use when acquiring unknown samples in PSD mode.