Chapter 5 Examining Spectrum Data

5-32

Applied Biosystems

5

6.

Specify reference masses to add by doing either of the
following:

Click a mass, then click OK.

Type new reference mass information in the
Name, Theoretical m/z, Charge, and Elemental
Composition fields, then click OK.

NOTE: You must type a minus sign (–) preceding
the charge in the Charge text box for negative
charge states.

NOTE: You can save changes to the Calibration
Reference File by clicking Save. For more
information, see “Modifying a calibration reference
file” on page 5-20.

The Automatic Calibration Settings dialog box is
displayed again (see Figure 5-7 on page 5-30).

7.

Repeat step 5 and step 6 to add all needed masses.

8.

To delete masses, do either of the following:

Select a reference, then click Delete Selected
Reference

Click Delete All References

9.

Enter Reference Matching Criteria:

Minimum Intensity—Peaks must be above this
intensity to be considered a match. Select the unit
for Minimum Intensity, % Relative Intensity or
Absolute Counts.

Mass Tolerance—Peaks must be within this
tolerance of the theoretical m/z to be considered a
match. Select the unit, m/z or ppm.