Chapter 5 Examining Spectrum Data

5-54

Applied Biosystems

5

General

guidelines for

setting

parameters

Refer to the following table to determine how to set Advance
Baseline Correction parameters and obtain the desired
baseline correction.

Condition

Set Parameters To

Baseline follows a rising peak cluster

Peak Width—Average peak width

Flexibility—0.5

Degree—0.5

Baseline is gently curving

Peak Width—10 to 20 times peak width

Flexibility—0

Degree—0.5 (decrease further to raise
the baseline)

Baseline rise should be ignored and
treated as signal

Peak Width—Half of the width of the
cluster to correct (in the illustration
shown, if the width of the cluster is 200,
set Peak Width to 100)

Flexibility—0

Degree—0.5 (decrease further to raise
the baseline)

Peak cluster