Peak Detection and Labeling Troubleshooting

Data Explorer

Software User’s Guide

9-19

9

Known isotope not labeled
with charge state

Charge State peak labels
disabled

Turn on Charge State
peak labels. See
Section 3.5.2, Setting
Chromatogram
and Spectrum Peak
Labels
.

Max Charge State
parameter set too low.
See example in “Max
Charge State set too low”
on page 3-33.

Set Max Charge State
correctly. See “Peak
Processing parameters
(spectrum data only)” on
page 3-26.

Peak List Filtering is
enabled, with Charge
State filter enabled and
set too low

Disable Peak List
Filtering. See “Filtering
the spectrum peak list” on
page 3-42.

Minimum Intensity set too
high to detect other
isotope peaks. See
example in “Effect of
Minimum Intensity” on
page 3-36.

Set Minimum Intensity
correctly. See “Advanced
Settings (spectrum data
only)” on page 3-28.

Filter width set too high to
determine other isotope
peaks

Set Filter Width correctly.
See “Advanced Settings
(spectrum data only)” on
page 3-28.

Table 9-12 Charge State and Isotope Determination Troubleshooting—Mariner

Only (Continued)

Symptom

Possible Cause

Action