Mass Deconvolution (Mariner Data Only)

Data Explorer

Software User’s Guide

5-41

5

5.

Type values for the following masses for the generated
zero-charge spectrum:

Center—Center mass

Half Width—Mass (±) from the Center mass
to include in the spectrum

Increment—Mass increment at which to
perform the calculation (0.1 for resolved
isotope peaks, 1.0 for unresolved isotope
peaks, >1.0 for noisy trace)

Figure 5-12 Zero-Charge Spectrum Conversion

Dialog Box

6.

Select or type the mass of the adduct ion to use in the
calculation. The mass of a proton (H

+

) is selected by

default.

7.

Click OK.

The deconvoluted trace (with a DECONV trace label)
replaces the original trace.

8.

To return to the original trace, see “Returning to the
original spectrum” on page 5-3.