HP Data Explorer 4 Series User Manual

Page 145

Advertising
background image

Deisotoping a Spectrum

Data Explorer

Software User’s Guide

3-47

3

Figure 3-17 Interpreting a Deisotoped Trace

Requirements

The Deisotope function requires a singly charged spectrum. If
the spectrum you are examining includes multiply charged
peaks, use the Single-Charge Conversion function on the
Process menu before using the Deisotope function. For
information, see Section 5.10, Converting to a Singly Charged
Spectrum (Mariner Data Only)
.

CAUTION

If you use the Deisotope function on multiply charged
peaks, invalid results are reported.

If first three peaks are part of

the same isotope cluster

If first three peaks are part of

the same isotope cluster, but

Contribution to expected
isotope ratio

contaminant is also present

If two isotope clusters

are present

Increased amplitude of first
peak indicates it is a
monoisotopic peak

Increased amplitude of first
peak indicates it is a
monoisotopic peak

Peak that is not
part of cluster

Contaminant

Increased amplitude of peaks
indicate they are
monoisotopic peaks

by contaminant

Peaks in original trace

Deisotoped trace

Peak that is not
part of cluster