Labmaster 9 zi-a – Teledyne LeCroy LabMaster 9Zi-A User Manual
Page 106
LabMaster 9 Zi-A
98
922157-00 Rev A
When DBI is selected, trace descriptor labels correspond with your choices.
So, a 30 or 36 GHz example where you decided to interleave both C1/C2
and C3/C4 pairs, trace descriptor labels are shown as follows:
C1 - Shown as inactive
C2 - Shown as active
C3 - Shown as active
C4 - Shown as inactive
M
ASTER
A
CQUISITION
M
ODULE
DBI
S
ETTINGS
When you use DBI and make correct connections and channel selections
from the TimeBase → Horizontal Setup... dialog, the system makes value
changes for you.
View these changes by touching the trace descriptor label for an
interleaved channel.
The Vertical Scale B Scale V/div (Timebase) value is automatically
set to 50 mV/div. If you set this to a different value and deactivate
DBI, the system retains the last one used when DBI is reactivated.
The channel LEDs form the following configuration in 30 or 36 GHz
mode on a Master Acquisition Module:
Channel 1
Channel 2
Channel 3
Channel 4
OFF, No LED
Light
▲
Green Down
Arrow
▼
Green Down
Arrow
▼
OFF, No LED
Light
▲
The channel LEDs form the following configuration in 45 GHz mode
on a Master Acquisition Module:
Channel 1
Channel 2
Channel 3
Channel 4
OFF, No LED
Light
▲
OFF, No LED
Light
▲
Green Down
Arrow
▼
OFF, No LED
Light
▲