Getting Started Manual

922157-00 Rev A

31

Gently pull the green tab on the cable plug to remove it.

Repeat the previous steps for every additional Slave Acquisition Module in
your system.

PLEASE NOTE THE FOLLOWING:

PCIe 4 Lane Cable plugs are keyed with a single groove along one wide
side of the plug. The plug must be inserted into the socket with the
groove aligned properly.

Be sure to connect from your Slave Acquisition Modules into the correct
channel groupings on the 9CZi-A Master Control Module; meaning, your
first Slave Acquisition Module is connected into DATALINK CHANNEL 1-4
input on the Master Control Module, second into the DATALINK CHANNEL
5-9 input, third into the DATALINK CHANNEL 9-12 input, fourth into the
DATALINK CHANNEL 13-16 input, and last into the DATALINK CHANNEL
17-20 input.

If you are connecting less than 5 of the Slave Acquisition Modules, the
PCIe 4 Lane Data Inputs on the back of the Master Control Modulemay
not be skipped and must be connected in consecutive order from the PCIe
4 Lane Data Outputs on the back of the corresponding slaves.