Standard Vertical Parameter Help Markers

Turning On Markers

1.

Touch Measure → Measure Setup... on the menu bar.

2.

Select mode Std Vertical, Std Horizontal, or My Measure.

3.

Touch the Show All button (as follows) to display Help Markers for
every parameter being measured on the displayed waveform (C2 in
the previous examples).

4.

Touch a Px tab for any listed parameter; it doesn't matter which.

5.

On the Parameter dialog, touch Help Markers and select either:

Simple - produces cursors and Measure Gate posts. The gate
posts are independently placeable for each parameter.

Detailed - produces cursors, Measure Gate posts, a label
identifying the parameter being measured, and a level
indicator and hysteresis band for "at level" parameters.

NOTE

: The choice of Simple or Detailed is applied to all parameters at the

same time. That is, if you choose Simple markers for one parameter, all
parameters are then displayed in this mode.

6.

Touch the Always On checkbox if you want to continuously display
Help Markers for this parameter.