Teledyne LeCroy LabMaster 9Zi-A User Manual

Page 189

Advertising
background image

Getting Started Manual

922157-00 Rev A

181

1. Touch Analysis → Pass/Fail Setup... from the menu bar.

2. Touch one of the Qx tabs and a setup dialog for the position is

shown.

3. Touch inside the Condition field in the main dialog and select

DualParamCompare.

4. Touch inside the Source1 and Source2 fields and select a source

from the pop-up menu.

5. Touch inside the ParamCompare mini-dialog field and select a

source from the pop-up menu.

6. Touch inside the Compare Values field and select All or Any from

the pop-up menu.

By selecting All, the test are true only if every waveform falls within
the set limit. Selecting Any makes the test true if just one waveform
falls within the limit.

7. Touch inside the Condition field on the ParamCompare part of the

dialog and select a math operator from the pop-up menu.

8. Touch inside the Limit field and enter a value, using the pop-up

numeric keypad. This value takes the dimension of the parameter
that you are testing. For example, if you are testing a time
parameter, the unit is seconds.

9. If you choose either WithinDeltaPct or WithinDeltaAbs from the

Condition menu respectively, touch inside the Delta field and enter
a value.