Prolink interface adapters – Teledyne LeCroy LabMaster 9Zi-A User Manual

Page 75

Advertising
background image

Getting Started Manual

922157-00 Rev A

67

NOTE:

The mounted connectors for the 2.92 mm and 2.4 mm interfaces look

identical, and the connector savers look very similar, so be sure to attach the
correct connector saver to achieve the desired bandwidth. The 2.92 mm connector
saver will fit over the 2.4 mm mounted connector (and vice versa) but will not yield
the full 45 GHz bandwidth.

ProLink Interface Adapters

Teledyne LeCroy's ProLink Adapters (LPA) provides the ability to connect a
cable to your oscilloscope that is either BNC, SMA, or K (2.92 mm)
terminated. These adapters are only for cable connection since they do not
provide the power and communication interfaces required for probe usage.
In fact, the absence of these connections on the adapters means that a
probe cannot be connected to the oscilloscope when one of these adapters
is connected.

The following cable connections are supported:

BNC, using the LPA-BNC Blind Mate Adapter (BMA) to BNC interface
adapter. This is not normally provide with your oscilloscope, but may
be ordered as an accessory.

SMA , using the LPA-SMA BMA-to-SMA adapter

K-Type (2.92 mm), using the LPA-K BMA-to-2.92 mm adapter (4 are
sent standard in 13 - 45 GHz units).

BMA-to-SMA, BMA-to-BNC, BMA-to-2.92 mm ProLink Adapters.

Advertising