LabMaster 9 Zi-A Channel Setup

You can set up to 20 channels on the Channel Setup dialog. You can make
adjustments to both the sensitivity and the positioning of the waveforms.

Setting up your channels

1. Press the appropriate channel push button (channel 1, for example).

OR
Touch Vertical → Channel Setup... from the menu bar.

2. Touch inside the checkbox next to a channel to display the trace.

NOTE:

Whenever you turn on a channel, a trace descriptor label is

displayed for that channel.

3. To adjust sensitivity and position, turn the V

ERTICAL

G

AIN

front panel

knob for the selected channel.
OR
Touch inside the Volts/Div field next to the channel and enter a
value. Click the keypad button to enter a value using the pop-up
keypad, or use the up/down arrows.

NOTE:

The set voltage is shown on the trace descriptor label and in the

Volts/Div field in the dialog.

4. To set the offset, turn the V

ERTICAL

O

FFSET

adjust front panel knob

directly above the channel button whose waveform you want to
move vertically. Or, you can touch inside the Offset field next to the
channel and type in a value on the pop-up keypad. To set the vertical
offset to zero, touch the Zero Offset button directly below the
Offset field.

5. To change the color assigned to a channel, click on the color box and

choose a color.

NOTE:

Front panel knobs can be used to adjust Vertical Offset and Gain for

channels 5-20, provided the trace descriptor label is active for the
corresponding channel.