Right-hand dialogs – Teledyne LeCroy LabMaster 9Zi-A User Manual

Page 199

Advertising
background image

Getting Started Manual

922157-00 Rev A

191

Right-Hand Dialogs

On the Test right-hand dialog, make a selection in the Test is True
when
group of buttons (as follows).

NOTE:

The previous selection is for

All In

and means the test is False

when a single waveform falls outside the mask.

From the Show Markers section of the dialog, choose whether or
not to have mask violations displayed by touching either the Off or
On button, respectively.

When loading a pre-existing mask, use the Load Mask right-hand
dialog. Using the File button, you can then enter the file name or
browse to its location.

When creating a new mask, use the Make Mask right-hand dialog.

Touch the Browse button to create a file name and location for the
mask if you want to save it.

Touch inside the Ver Delta and Hor Delta fields and enter boundary
values, using the pop-up numeric keypad. You can also use the Make
from Trace
button to instantly create a mask from a new waveform.

When creating a new mask, use the Make Mask right-hand dialog.

On the Gate right-hand dialog, you can enter values in the Start and
Stop fields to constrain the mask to a portion of the waveform. Or,
you can simply touch and drag the Gate posts, which initially are
placed at the extreme left and right ends of the grid.