LabMaster 9 Zi-A

124

922157-00 Rev A

If you perform ChannelSync Cal on multiple gain settings, perform
it first on the 100 mV/div setting, then perform it on other gain
settings (with “Reference Correction” values entered as
appropriate for those other gain settings.)

For the highest possible accuracy across the full input range of the
oscilloscope, a ChannelSync Cal should be performed on at least
one gain setting in each of the following gain ranges:

13-20 GHz ProLink

"A" Inputs

30-45 GHz 2.92mm or 2.4mm

"B" Inputs

<11.8 mV/div

10 mV/div

11.9 to 17.0 mV/div

10.1 to 12.6 mV/div

17.1 to 20.5 mV/div

12.7 to 15.8 mV/div

20.6 to 30.0 mV/div

15.9 to 20.0 mV/div

30.1 to 39.6 mV/div

20.1 to 25.0 mV/div

39.6 to 50.0 mV/div

25.1 to 31.5 mV/div

50.1 to 90.0 mV/div

31.6 to 39.5 mV/div

90.1 to 100.0 mV/div

39.6 to 50.0 mV/div

100.1 to 118.0 mV/div

50.1 to 63.0 mV/div

118.1 to 170.0 mV/div

63.1 to 79.0 mV/div

170.1 to 205.0 mV/div

79.1 to 82.0 mV/div

205.1 to 300.0 mV/div

82.1 to 100.0 mV/div

300.1 to 395.0 mV/div

100.1 to 126.0 mV/div

395.1 to 500.0 mV/div

126.1 to 158.0 mV/div

500.1 to 900.0 mV/div

158.1 to 200.0 mV/div

900.1 to 1000 mV/div

200.1 to 250.0 mV/div

If you add cables or probes to your setup subsequent to running
the ChannelSync Cal, you must manually deskew these items in the
Channel dialog to account for their propagation delays. The values
you provide on the Channel dialog (in the deskew section) are
additional to values shown on the ChannelSync Calibration dialog.