Vertical Offset - For each channel, you can select between zero
vertical offset or to set the offset to a specific value.

Coupling - The maximum input voltage depends on the input used.
Values are displayed on the front of the oscilloscope. Whenever the
voltage exceeds this limit, the coupling mode automatically switches
from DC 50 Ω to GROUND. You then have to manually reset the
coupling to DC 50 Ω.

CAUTION. While the unit does provide this protection,
damage can still occur if extreme voltages are applied

Bandwidth - Bandwidth filters are available at the following levels:

Full (all X-Stream oscilloscopes), 30 GHz, 25 GHz, 20 GHz, 16 GHz, 13
GHz, 8 GHz, 6 GHz, 4 GHz , 3 GHz , 1 GHz , 200 MHz (all X-Stream
oscilloscopes), and 20 MHz (all X-Stream oscilloscopes).

WaveMaster and LabMaster - 45 GHz, 30 GHz, 25 GHz, 20 GHz, 16
GHz, 13 GHz, 8 GHz, 6 GHz.

WavePro and WaveMaster - 4 GHz, 3 GHz, 1 GHz, 200 MHz, and 20
MHz.

WaveRunner - 1 GHz, 200 MHz, and 20 MHz.

PLEASE NOTE THE FOLLOWING:

Each oscilloscope contains only the Full selection and those bandwidths
less than the rated oscilloscope bandwidth.

Anti-aliasing filters are applied to limit bandwidth at lower sample rates,
regardless of bandwidth limit selection.

Probe - Teledyne LeCroy's ProBus, ProLink, and 2.92 mm system
automatically senses probes and sets their attenuation for you. For each
channel, the probe attenuation can also be set manually.