Source[1]:apply:sinusoid, Source[1]:apply:square – GW Instek AFG-2000 User Manual User Manual
Page 100
AFG-2000 Series User Manual
98
This means that the magnitude of the DC offset is
determined by the output amplitude.
If the specified DC offset is out of range, the
maximum/minimum offset will be set instead. A
―-222‖ error will be generated from the remote
terminal.
SOURce[1]:APPLy:SINusoid
Set
Description
Outputs a sine wave when the command has
executed. Frequency, amplitude and offset can also
be set.
Syntax
SOURce[1]:APPLy:SINusoid [<frequency>
[,<amplitude> [,<offset>] ]]
Parameter
<frequency>
0.1Hz~25MHz*
<amplitude>
1mV~10Vpp (50Ω)
<offset>
-5V ~ +5V (50Ω)
*AFG-2005/2105 limited to 5MHz, AFG-2012/2112
limited to 12MHz.
Example
SOURce1:APPL:SIN MAX, 3.0, -2.5
Outputs a 3Vpp sine wave at 25MHz (max
frequency) with a -2.5V offset.
SOURce[1]:APPLy:SQUare
Set
Description
Outputs a square wave when the command has
executed. Frequency, amplitude and offset can also
be set. The duty cycle is fixed to 50%.
Syntax
SOURce[1]:APPLy:SQUare [<frequency> [,<amplitude>
[,<offset>] ]]
Parameter
<frequency>
0.1Hz ~ 25MHz*
<
amplitude
>
1mV~10V (50Ω)