SOURce[1]:SWEep:SPACing

Set

Query

Description

Sets linear or logarithmic sweep spacing. The
default spacing is linear.

Syntax

SOURce[1]:SWEep:SPACing {LINear|LOGarithmic}

Example

SOUR1:SWE:SPAC LIN

Sets the spacing to linear.

Query Syntax

SOURce[1]:SWEep:SPACing?

Return Parameter LIN

Linear spacing

LOG

Logarithmic spacing

Query Example

SOUR1:SWE:SPAC?
>LIN

The spacing is currently set as linear.

SOURce[1]:SWEep:RATE

Set

Query

Description

Sets or queries the sweep rate. The default sweep
rate is 100 Hz. This command is the equivalent to
using the Rate function on the front panel.

Note

The function generator automatically determines
the number of frequency points that are used for
the sweep based on the sweep rate.

Syntax

SOURce[1]:SWEep:RATE
{<Hz>|MINimum|MAXimum}

Parameter

<Hz>

2mHz ~1kHz (equivalent to a sweep
time of 500s ~ 1ms)

Example

SOUR1:SWE:RATE +1.0000E+00

Sets the rate to 1 Hz (1 second).

Query Syntax

SOURce[1]:SWEep:RATE? {<Hz>|
MINimum|MAXimum}

Return Parameter <NR3>

Returns sweep rate in Hz.