Source[1]:dcoffset, Source[1]:square:dcycle – GW Instek AFG-2000 User Manual User Manual

Page 108

Advertising
background image

AFG-2000 Series User Manual

106

The maximum amplitude that can be set for the
current function is 5 volts.

SOURce[1]:DCOffset

Set

Query

Description

Sets or queries the DC offset for the current mode.

Note

The offset parameter can be set to MINimum or
MAXimum. The default offset is 0 volts. The offset
is limited by the output amplitude as shown
below.

|Voffset| < Vmax – Vpp/2

If the output specified is out of range, the
maximum offset will be set.
The maximum offset is ±5V into 50Ω).

Syntax

SOURce[1]:DCOffset {< offset>
|MINimum|MAXimum}

Example

SOUR1:DCO MAX

Sets the offset to the maximum for the current
mode.

Query Syntax

SOURce[1]:DCOffset? {MINimum|MAXimum}

Return Parameter <NR3>

Returns the offset for the current mode.

Query Example

SOUR1:DCO?

>+3.0000E+00

The offset for the current mode is set to +3 volts.

SOURce[1]:SQUare:DCYCle

Set

Query

Description

Sets or queries the duty cycle for square waves
only. The setting is remembered if the function
mode is changed. The default duty cycle is 50%.

Note

The duty cycle of square waveforms depend on the
frequency settings.