Source[1]:sweep:source – GW Instek AFG-2000 User Manual User Manual

Page 129

Advertising
background image

REMOTE INTERFACE

127

Query Example

SOUR1:SWE:RATE?
>+2.0000000E+01

Returns the sweep rate (20 Hz).

SOURce[1]:SWEep:SOURce

Set

Query

Description

Sets or queries the trigger source as immediate
(internal) or external. Immediate (internal) is the
default trigger source. IMMediate will constantly
output a swept waveform. EXTernal will output a
swept waveform after each external trigger pulse
(TTL positive edge).

Note

If EXTernal is selected, the trigger period must be
greater than the sweep time + 125nS.

Syntax

SOURce[1]: SWEep:SOURce {IMMediate|EXTernal }

Example

SOUR1: SWE:SOUR EXT

Sets the sweep source to external.

Query Syntax

SOURce[1]: SWEep:SOURce?

Return Parameter IMM

Immediate

EXT

External

Query Example

SOUR1:SWE:SOUR?
>IMM

The sweep source is set to immediate.