Afg-2000 series user manual – GW Instek AFG-2000 User Manual User Manual

Page 60

Advertising
background image

AFG-2000 Series User Manual

58

Connection
(EXT source only)

For external sources,
connect the modulation
source signal to the MOD
input port on the rear
panel.

OUTPUT

INPUT

MOD

Counter

Trigger

MOD

Note

When the source is set to EXT (external) the
carrier waveform is modulated by an external
signal. The frequency deviation is controlled by
the ±5V signal that is input into the MOD input
port. The ±5V input signal directly corresponds
to the set frequency deviation. +5V increases the
frequency by the set deviation frequency and -5V
reduces the frequency to below the carrier
frequency by the amount set by the deviation
frequency. For example: if the deviation
frequency is set to 1kHz, an input voltage of +5V
will increase the frequency to 1kHz, whilst an
input voltage of -5V will reduce the frequency
below that of the carrier by 1kHz.

Example:
Source = INT

Example: External
MOD input signal

FM output

MOD input

signal

0V

0V

+5V

-5V

Advertising