5. MEASUREMENTS

5.1. CONTINUITY - TWO WIRE METHOD (RPE-2WIRE)

Complying with EN 60204-1, continuity of protective bonding circuit between PE terminal and

relevant points of the protective conductor system must be checked by injecting a measurement

current of 0.2 A up to 10 A approx.

5.1.1. RPE-2WIRE DISPLAY EXPLANATION

Adjustable/selectable parameters:

Im

NOM

- nominal measurement current

200 mA or 25 A AC

LIMIT (meas. current 200 mA) - continuity limit value 0.01

 19.99, 20.0  200.0 

LIMIT (meas

.

current 25 A) - continuity limit mode

STANDARD, 60204 SET Z or 60204 SET L

LIMIT (test current 25 A, STANDARD mode) -
continuity limit value

0.01

 20.00 

MODE - measurement mode

MANUAL or TIMER

CAL (meas. current 200 mA) - calibration of test leads 0.00

 5.00 

CAL (meas

.

current 25 A) - calibration of test leads 0.000

 1.999, 2.00  5.00 

TIMER - measurement time

00:01

 60:00 (1 s  60 min), resolution 1 s

LENGTH - wire length

0.1

 999.9 m, resolution 0.1 m

SECTION - wire section

1, 1.5, 2.5, 4, 6, 10, 16, 25, 35, 50 or 70 mm

2

MATERIAL - wire material

Cu (Cupper) or Al (Aluminium)

Z

LINE

- input line impedance

0.001

 2.000 , resolution 0.001 

PROTECTION - over-current protection device

MCB B, MCB C, MCB D, MCB K, FUSE gG or

FUSE

aM

In - nominal current of protection device

Depends on selected protection device, see the

chapter 5.1.3. LIMIT VALUE ADJUSTMENT

Figure 5

: Display with R

PE-2WIRE

test result

1 ..... Selected function.

2 ..... Progress bar, it follows measurement time during the measurement (in TIMER mode only).
3 ..... Sub-result - measurement current Im that was flowing through UUT during the measurement.

4 ..... Measurement screen touch-screen key.
5 ..... Im NOM touch-screen key to select nominal test current (200 mA or 25 A). Currently selected value

is displayed on the bottom of the key.