T1 s – HT instruments FULLTEST3 User Manual

Page 47

Advertising
background image

FULLTEST

3

EN - 46

8) Test result will be displayed in green colour equipped with green

symbol and with beep-beep

sound if it is lower than or equal to 60 VRMS (may be AC or DC, see the unit). If the result is higher
than 60 VRMS, then it will be displayed

in red colour equipped with red

symbol and with longer

beep sound

. See the display outlook with test result on the

figure 32.

9)

Save the test result by pressing the SAVE hard key twice,

for further instructions see the

“MEMORIZING EXAMPLE” section.

10)

Press the EXIT hard key to clear displayed result, actual input voltage U

IN

and actual trigger

voltage U

TRIG

will start to be displayed again.

PLUG/1s mode is selected:
4) Select measurement screen by pressing the

(3) touch-screen key and check all settings again.

5) Connect test leads according to the figure below.

t

1 s

RES

<

Figure 35

: Connection of test leads in U

RES

PLUG measurement

6) READY, DISCONNECT UUT message will appear when input voltage U

IN

within required range

100

 460 VAC is present. Carry out the measurement by disconnecting UUT. Disconnection of

UUT according to the figure 35 means disconnection of multiple plug adapter.

7)

Test result will be displayed in green colour equipped with green

symbol and with beep-beep

sound if it is lower than or equal to 60 VRMS (may be AC or DC, see the unit). If the result is higher
than 60 VRMS, then it will be displayed

in red colour equipped with red

symbol and with longer

beep sound

. See the display outlook with test result on the

figure 32.

8)

Save the test result by pressing the SAVE hard key twice,

for further instructions see the

“MEMORIZING EXAMPLE” section.

9)

Press the EXIT hard key to clear displayed result, actual input voltage U

IN

will start to be displayed

again.

CAUTION

Do not use the START button in this function, it has

no function!