HT instruments FULLTEST3 User Manual

Page 27

Advertising
background image

FULLTEST

3

EN - 26

5.4.2. DIELECTRIC DISPLAY EXPLANATION

Adjustable/selectable parameters:
MODE - operation mode

MANUAL, RAMP or BURN

U

TEST NOM -

nominal test voltage

250

 5100 V AC

LIMIT - current limit

1

 110 mA

CHAR - current character

I

APP

or I

REAL

Figure 13

: Display with DIELECTRIC test result

1 ..... Selected function.

2 ..... Progress bar, it follows test time during the measurement (in RAMP mode only).
3 ..... Test voltage applied during the measurement.

4 ..... Measurement screen touch-screen key.
5 ..... MODE touch-screen key to select operation mode (MANUAL, RAMP or BURN). Currently selected

mode is displayed on the bottom of the key.

6 ..... U

TEST NOM

touch-screen key to select nominal test voltage (250 up to 5100 V

). Currently

selected value is displayed on the bottom of the key.

7 ..... LIMIT touch-screen key. Currently selected limit leakage current (trip out current) is displayed on

the bottom of the key.

8 ..... CHAR (character) touch-screen key to select the character of displayed leakage current (I

APP

or

I

REAL

). Currently selected character is displayed on the bottom of the key.

9 ..... Leakage current in green colour if the result is lower than or equal to set limit value. If break-

through occurred during the test then limit value will be displayed in red colour.

10 ... Real time clock (hh.mm.ss).
11 ... Unit of the test result (mA).

12 ... Set measurement time (in RAMP mode only).
13 ... Measurement result status (symbol

in green colour - result lower than or equal to set limit

value, symbol

in red colour - break through occurred during the test or result higher than set

limit value).