Insulation resistance (M

)

DC test voltage

(V)

Display range

(M

)

Resolution

(M

)

Accuracy

Overvoltage

protection

100

0.00

 9.99

0.01

± (3% rdg. + 3 dgt)

CAT III 300 V

10.0

 20.0

0.1

20.0

 99.9

0.1

± 5% rdg.

250

0.00

 9.99

0.01

± (3% rdg. + 3 dgt)

10.0

 20.0

0.1

20.0

 99.9

0.1

± 5% rdg.

100

 250

1

500

0.00

 9.99

0.01

± (3% rdg. + 3 dgt)

10.0

 20.0

0.1

20.0

 99.9

0.1

± 5% rdg.

100

 500

1

1000

0.00

 9.99

0.01

± (3% rdg. + 3 dgt)

10.0

 20.0

0.1

20.0

 99.9

0.1

± 5% rdg.

100

 1000

1

Test voltage tolerance:

(-0%

 +25%) of U

N

Test current:

> 1 mA (up to U

N

/1mA)

Short-circuit current:

< 15 mA

Discharge:

Internal resistance of 2 M

 (after finishing the measurement)

Detection of external voltage

Yes

-

U

EXT

lim = 10 V AC (between R

ISO

+ and

R

ISO

- terminals

before measurement)

-

U

EXT

lim = 50 V AC (between R

ISO

+ and

R

ISO

- terminals

during measurement)

- U

EXT

lim = 50 V AC approx. (between any R

ISO

terminal

and GND)

- U

EXT

lim = -10 V DC (between R

ISO

+ and

R

ISO

- terminals

during measurement)

Advertising