HT instruments FULLTEST3 User Manual

Page 44

Advertising
background image

FULLTEST

3

EN - 43

To include standard mains over-voltage, the measured residual voltage is scaled to peak value of max.
possible mains over-voltage, i.e.:

a) Selected nominal voltage Un = 230 V
Up = 230 V Ч 1.1 Ч 1.41 = 358 V ....................... system voltage 230 V is recognized

Up = 400 V Ч 1.1 Ч 1.41 = 620 V ....................... system voltage 400 V is recognized

b) Selected nominal voltage Un = 240 V

Up = 240 V Ч 1.1 Ч 1.41 = 372 V ....................... system voltage 240 V is recognized
Up = 415 V Ч 1.1 Ч 1.41 = 644 V ....................... system voltage 415 V is recognized

If actual mains voltage differs from nominal system voltage more than

10%, the FULLTEST 3 scales

the result to peak value of actual input voltage.

Example 1 (Un = 230 V):

U

IN

= 173 V (the value differs more than 10% from 230 V), result is scaled to 173 V × 1.41 = 244 V


Example 2 (Un = 230 V):

U

IN

= 209 V (the value differs less than 10% from 230 V), result is scaled to 230 V Ч 1.1 Ч 1.41 = 358 V

5.8.2. EXPLANATION OF NONLINEAR MODE

In non-linear mode it is assumed there are also

"

non-linear

"

or unknown components involved in

discharge process (relays, gas lamps etc.) and therefore discharge characteristic is non-exponential or it

is unpredictable, see the diagram below.

t

1s / 5s

U

Some relay may be switched off

U

(measured)

RES

Coincidential switch off

at any mains voltage

Figure 31

: Discharge diagram in non-linear circumstances

In this case result can not be scaled to peak value, so it must be assured that switch off occurs at max.
input voltage i.e. at pick value (± 5%), otherwise measured result is not relevant. Measured value is

then registered and evaluated.

Advertising