HT instruments FULLTEST3 User Manual

Page 87

Advertising
background image

FULLTEST

3

EN - 86

TRMS Load Current (POWER)

Display range

A

Resolution

A

Accuracy

Overvoltage protection

0.00

 0.99

0.01

± (3% rdg. + 3 dgt)

CAT II 300 V

1.0

 20.0

0.1

Frequency range:

15

 723 Hz

Over-range limit:

Yes, the measurement will be automatically interrupted 10 s
after exceeding 16 A

Pre-Test

Grounding of PE terminal on wall schuko socket

TRMS Mains Voltage (POWER)

Display range

V

Resolution

V

Accuracy

Overvoltage protection

195

 253

1

± (2% rdg. + 2 dgt)

CAT II 300 V

Frequency range:

15

 723 Hz

Apparent Power / Active Power (POWER)

Display range

VA/W

Resolution

VA/W

Accuracy

Overvoltage protection

0.0

 99.9

0.1

± (5% rdg. + 10 dgt)

CAT II 300 V

100

 999

1

± (5% rdg. + 3 dgt)

1.00

 5.06 k

10

Limit value P

APP:

Adjustable

6

 999 VA, 1.00 kVA  5.06 kVA

Power factor (POWER)

Display range

Resolution

Accuracy

Overvoltage protection

0.00

 1.00

0.01

Respect the accuracy of P

APP

and P

ACT

CAT II 300 V

TRMS Leakage Current I

PE

(POWER)

Display range

Resolution

Accuracy

Overvoltage protection

0.25

 19.99 mA

0.01 mA

± (3% rdg. + 3 dgt)

CAT II 300 V

20.0

 49.9 mA

0.1 mA

0.05

 0.99 A

0.01 A

1.0

 10.0 A

0.1 A

Load current influence:

± 0.01 mA / A

Limit value:

Adjustable 0.25 mA

 10.00 A, standard limit value 3.50 mA

Frequency range:

40 Hz

 100 kHz (characteristics according to

DIN EN 61010-1 Annex A, fig. A.1 and VDE 0411 Part 1)

Measurement method:

Differential

Mains cord polarity exchange

Internally by using the L POS key

UUT mains on switch test:

Yes, the test done by measuring L current, limit value 25 mA

Over-range limit (I

PE

):

Yes, the measurement will be automatically interrupted 10 s

after exceeding 10 A

Over-range limit (I

L

):

Yes, the measurement will be automatically interrupted 10 s

after exceeding 18 A