HT instruments FULLTEST3 User Manual

Page 56

Advertising
background image

FULLTEST

3

EN - 55

5.12.

LEAKAGE CURRENT (ILEAK)

5.12.1. LEAKAGE DISPLAY EXPLANATION

Adjustable/selectable parameters:

MODE - measurement mode

CLAMP or SOCKET

LIMIT - leakage current limit value 0.1

 100.0 mA, 101  1000 mA (CLAMP mode, range 1000 mA)

0.1

 100.0 A (CLAMP mode, range 100.0 A)

1

 1000 A (CLAMP mode, range 1000 A)

0.01

 19.99 mA, 20.0  49.9 mA, 0.05  0.99 A, 1.0  10.0 A

(SOCKET

mode)

RANGE - clamp measurement range 1000 mA, 100.0 A or 1000 A

L POS - phase position

LEFT or RIGHT

Figure 42

: Display with I

LEAK

test result in CLAMP mode

1 ..... Selected function.
2 ..... Measurement screen touch-screen key.

3 ..... MODE touch-screen key to select measurement mode (SOCKET or CLAMP). Currently selected

mode is displayed on the bottom of the key.

4 ..... RANGE touch-screen key to select the CLAMP measurement range. Currently selected range is

displayed on the bottom of the key.

5 ..... LIMIT touch-screen key to select limit leakage current. Currently selected value is displayed on the

bottom of the key.

6 ..... Measurement result (in green colour - result OK, in red colour - result not OK).
7 ..... Real time clock (hh.mm.ss).

8 ..... Unit of the test result.
9 ..... Measurement result status (symbol

in green colour - result OK, symbol

in red colour - result

not OK).