HT instruments FULLTEST3 User Manual

Page 82

Advertising
background image

FULLTEST

3

EN - 81

Dielectric withstanding test (DIELECTRIC)

Nominal test

voltage U

N

(V)

Output

Resolution

(V)

Accuracy of

output voltage

Overvoltage

protection

250

 800

COM & 0.25

0.80 kV

10 ±

3%

U

N

CAT III 300 V

810

 2500

COM & 0.81

2.50 kV

2510

 5100

COM & 2.51

5.10 kV

Nominal test voltage U

N

: Adjustable

250

 5100 V, 50/60 Hz (floating) in steps of 10 V

Distortion of test voltage:

Crest factor = 1.414 ± 5%

Measurement modes:

MANUAL, RAMP (timer) or BURN

Output power:

500 VA at 5100 V

Leakage current I

APP

:

Display range

(mA)

Resolution

(mA)

Accuracy

0

 200

1

± (3% rdg + 2 mA)


Leakage current I

REAL

:

Display range

(mA)

Resolution

(mA)

Accuracy

0

 110

1

± (3% rdg + 4 mA)

Nominal breaking current (I

APP

or I

REAL

): Adjustable 1

 110 mA in steps of 1 mA

Short-circuit current:

> 200 mA

Breaking time:

< 30 ms


Note!

Always connect COM terminal to GND if measured OUT is grounded, otherwise possible capacitive
leakage current may flow to ground which may disturb the measurement!