HT instruments FULLTEST3 User Manual

Page 45

Advertising
background image

FULLTEST

3

EN - 44

5.8.3. U

RES

DISPLAY EXPLANATION

Selectable parameters:
MODE - measurement mode LINEAR or NONLINEAR

CON - connection

INT (measurement on internal components) or PLUG (measurement on
1P/3P plug)

LIMIT t - limit time

1 s or 5 s

Figure 32

: Display with U

RES

test result

1 ..... Selected function.

2 ..... Input voltage U

IN

and trigger voltage U

TRIG

.

3 ..... Measurement screen touch-screen key.

4 ..... MODE touch-screen key to select measurement mode (LINEAR or NONLINEAR). Currently

selected mode is displayed on the bottom of the key.

5 ..... CON (connection) touch-screen key to select measurement connection (INT or PLUG). Currently

selected connection is displayed on the bottom of the key.

6 ..... LIMIT t touch-screen key to select limit time (1 s or 5 s), valid for internal measurement only.

Currently selected limit value is displayed on the bottom of the key.

7 ..... Measurement result (in green colour - result OK, in red colour - result not OK).
8 ..... Real time clock (hh.mm.ss).

9 ..... Unit of the test result. As measured U

RES

voltage may be alternating or direct there is an

appropriate information AC or DC added to the unit.

10 ... Measurement result status (symbol

in green colour - result OK, symbol

in red colour - result

not OK).

5.8.4. TRIGGER CONDITIONS

The instrument recognizes disconnection of mains voltage on TRIG input (INT measurement) or on
U

RES

input (PLUG measurement) when one of the following two conditions occurs:

 If mean value of rectified input voltage drops down with a slope of at least 25 V / s (mean value

measured each period), then trigger is activated and the measurement starts to run.
This condition will occur for example if AC or DC input voltage starts to decrease.

 Momentary value of current half period is compared with momentary value of previous half period

(the same polarity). If there is a difference higher than 10%, then trigger is activated and the

measurement starts to run.

This condition will occur for example if AC voltage changes to DC.

The above two conditions are active on U

RES

input in PLUG mode and on U

TRIG

input in INT mode.