HT instruments FULLTEST3 User Manual

Page 46

Advertising
background image

FULLTEST

3

EN - 45

5.8.5. U

RES

MEASUREMENT

Measured quantities and display ranges:

Residual voltage on power plug U

RES

10

 460 V AC or 10  650 V DC

Residual voltage on internal components U

RES

10

 460 V AC or 10  650 V DC

Input requirements:
Input voltage U

IN

(READY condition in PLUG mode)

43

 460 V AC

Trigger voltage U

TRIG

(READY condition in INT mode) 100

 460 V AC

1) Select U

RES

function by pressing the FUNC key first.

2) Check selected mode and modify it if needed by pressing the MODE (4) touch-screen key first.

3) Check selected connection and modify it if needed by pressing the CON (5) touch-screen key first.

INT 4-WIRE mode is selected:
4) Check selected limit time and modify it if needed by pressing the LIMIT t (6) touch-screen key first.

5) Select measurement screen by pressing the

(3) touch-screen key and check all settings again.

6) Connect test leads according to one of the figures below.

Figure 33

: Connection of test leads in U

RES

INT measurement on 1P/3P plugged machines

Figure 34

: Connection of test leads in U

RES

INT measurement on fixed connected machines

7) READY, DISCONNECT UUT message will appear when U

TRIG

voltage within required range 100

 460 VAC is present. Carry out the measurement by disconnecting UUT. Disconnection of UUT
according to the figure 33 means disconnection of multiple plug adapter. Disconnection of UUT
according to the figure 34 means turning off mains switch.