HT instruments FULLTEST3 User Manual

Page 22

Advertising
background image

FULLTEST

3

EN - 21

with beep-beep sound (result OK) or with

red

symbol and with longer beep sound (result not

OK)

. See the display outlook with test result on

Figure 8.

7)

Save the test result by pressing the SAVE hard key twice,

for further instructions see the

“MEMORIZING EXAMPLE” section.

CAUTION

Max external voltage between two R

PE

or between two SENSE test terminals

is 10 VAC, no DC external voltage is allowed! In case of higher external

voltage fuse F4 (

T20A/500V, 6.3×32 mm

) may blow!

If SENSE test leads are not connected, then measurement result will include

also the resistance of current test leads.

Measurement time in MANUAL mode is limited to 60 min!

The following specific information can be shown on the display during measurement:

Information displayed

Description

EXTERNAL VOLTAGE

 External voltage higher than 3 VAC is applied between

two R

PE

or between two SENSE test terminals

(measurement is not running) or higher than 10 VAC

(measurement is running).

 External voltage higher than 5 30 V is applied between

any R

PE

or SENSE test terminal and GND.

Remove external voltage!

FUSE F4!

Fuse F4 is blown.

EROR1!

Internal fuse may be blown!
The fuse is not customer replaceable, send the tester into

a service department.