In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual

Page 15

Advertising
background image

Name

Command

Response

Start Measurement

Start Measurement CRC


















Send Data







Additional Data

aM!

aMC!


















aD0!







aD1!...aD9!

a004n<CR><LF>
n parameters will be available for reading by the Send
Data command within 4 seconds. A service request
(a<CR><LF>) will be sent when the parameters are
ready. The number of parameters returned is
determined by the SDI-12 configuration file. The default
value for n is 7. The default parameters are:

1: DO concentration in mg/L
2: DO percent saturation

3: Temperature in

°C

4: Specific conductivity in uS/cm
5: Depth in feet
6: pH
7: ORP in millivolts

If the device is configured to output more than 9
parameters, only the first 9 parameters can be
measured by this command.


a<values><CR><LF> or
a<values><CRC><CR><LF>
The number and type of parameters returned is
determined by the SDI-12 configuration file. At most 3
parameters are returned in a Send Data command. If
more than 3 parameters are available, they are
returned using the additional data command.

a<values><CR><LF> or
a<values><CRC><CR><LF>
At most 3 parameters are returned per additional data
request.

Additional Measurements

Additional with CRC


Send Data


Additional Data

aM1! … aM9!

aMC1!… aMC9!


aD0!


aD1! … aD9!

a0000<CR><LF>

No additional measurements shall be started by the
device.

a<CR><LF> or
a<CRC><CR><LF>

No values are returned after an Additional
Measurement command.