4.9 Calibration Registers—RDO

Values in the calibration registers determine how the sensor parameters are calculated.

Register

Size

(registers)

Mode & Access Level

(R/W)

Data

Type

Description

0118

2

R1/W3

float

Live salinity value (PSU)

0120

2

R1/W3

float

Default salinity value
(PSU, default = 0.0)

0122

2

R1/W3

float

Live barometric pressure (P

B

, mbar)

0124

2

R1/W3

float

Default barometric pressure

(mbar, default = 1013.25)

0126

2

R1/W3

float

100% saturation calibration

reading (mg/L)

0128

2

R1/W3

float

100% saturation temperature

reading (°C)

0130

2

R1/W3

float

100% saturation salinity value

(PSU)

0132

2

R1/W3

float

100% saturation

barometric pressure (mbar)

0134

2

R1/W3

float

0% saturation calibration

reading (mg/L)

0136

2

R1/W3

float

0% saturation temperature

reading (°C)

0138

2

R1/W3

float

Calibration slope (default = 1.0)

0140

2

R1/W3

float

Calibration offset (default = 0.0)