In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual

Page 22

Advertising
background image

Name

Command

Response

Set pH Slope










Send Data

aXHSpd.d!










aD0!

a0011<CR><LF>
One result is available in 1 second for reading by
the Send Data command. A service request
(a<CR><LF>) will be sent when the result is ready.

pd.d = pH slope (pH/mV at 25 °C, must be less
than zero)
p = polarity sign (+ or -)
d = digits (1 to 7)
. = decimal point (optional)

a+s<CR><LF>
s = command status, 1 = command successful, 0 =
command failed (sensor not plugged in).

Read pH Slope



Send Data

aXHS!



aD0!

a0011<CR><LF>
One result is available in 1 second for reading by
the Send Data command.

a+pd.d<CR><LF>
pd.d = pH slope (pH/mV at 25°C, returns -9999 if
sensor is not plugged in)

Set pH Offset










Send Data

aXHOpd.d!










aD0!

a0011<CR><LF>
One result is available in 1 second for reading by
the Send Data command. A service request
(a<CR><LF>) will be sent when the result is ready.

pd.d = pH offset (must be greater than zero)
p = polarity sign (+ or -)
d = digits (1 to 7)
. = decimal point (optional)

a+s<CR><LF>
s = command status, 1 = command successful, 0 =
command failed (sensor not plugged in)

Read pH Offset



Send Data

aXHO!



aD0!

a0011<CR><LF>
One result is available in 1 second for reading by
the Send Data command.

a+pd.d<CR><LF>
pd.d = pH offset (pH, returns -9999 if sensor is not
plugged in)

22