15 probe common registers, Probe common registers – In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual

Page 69

Advertising
background image

All registers in this document are 1-based. This means the actual packets sent to the devices
must have a data address 1 less than what the register number is in this document.

8.15 Probe Common Registers

Register

XXXX

Size

(registers

Mode

(R/W)

Data

Type

Description

9000

1

R

ushort

Register Map Template Version (SIS Level)

9001

1

R/W

ushort

Device ID

9002-003

2

R/W

ulong

Device Serial Number

9004-006

3

R/W

time

Manufacture Date

9007

1

R

ushort

Firmware version * 100

9008

1

R

ushort

Boot Code version * 100

9009

1

R

ushort

Hardware version

9010

1

R

ushort

Max Data Logs

9011-012

2

R

ulong

Total Data Log Memory (bytes)

9013-014

2

R

ulong

Total Battery Ticks

9015-017

3

R/W

time

Last Battery Change

9018

1

Reserved

9019-050

32

R/W

string

Device Name

9051-082

32

R/W

string

Site Name

9083-086

4

R/W

double

Latitude Coordinate (degrees) (negative = east)

9087-090

4

R/W

double

Longitude Coordinate (degrees) (negative = south)

9091-094

4

R/W

double

Altitude Coordinate (meters)

9095-096

2

Reserved

9097-099

3

R/W

time

Current time (UTC)

9100-101

2

R

32 bits

Device Status

9102-103

2

R/W

ulong

Used Battery Ticks

9104-105

2

R

ulong

Used Data Log Memory (bytes)

69