In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual

Page 21

Advertising
background image

Name

Command

Response

Restore Level
Factory Calibration




Send Data

aXLFC!





aD0!

a0011<CR><LF>
One result is available in 1 second for reading by
the Send Data command. Restores pressure and
level calibrations to their factory defaults. No
change to other parameter settings.

a+s<CR><LF>
s = command status, 1 = command successful, 0 =
command failed.

Zero Pressure




Send Data

aXPZ!




aD0!

a0011<CR><LF>
One result is available in 1 second for reading by
the Send Data command. A service request
(a<CR><LF>) will be sent when the result is ready.

a+s<CR><LF>
s = command status, 1 = command successful, 0 =
command failed.

Set Level Reference










Send Data

aXLRpd.d!










aD0!

a0011<CR><LF>
One result is available in 1 second for reading by
the Send Data command. A service request
(a<CR><LF>) will be sent when the result is ready.

pd.d = level reference in current units
p = polarity sign (+ or -)
d = digits (1 to 7)
. = decimal point (optional)

a+s<CR><LF>
s = command status, 1 = command successful, 0 =
command failed.

Restore pH/ORP Factory
Calibration



Send Data

aXHFC!




aD0!

a0011<CR><LF>
One result is available in 1 second for reading by
the Send Data command. Restores pH and ORP
calibrations to their factory defaults.

a+s<CR><LF>
s = command status, 1 = command successful, 0 =
command failed.

21

Advertising
This manual is related to the following products: