2. SDI-12 Interface

The Multi-PRO 400 Instrument adheres to the Serial Digital Interface Standard for
Microprocessor-based sensors, version 1.3, and the extensions to the specifications identified in
this document. This section identifies the device-specific implementation of the standard. For
more information on SDI-12 see

www.sdi-12.org

.

2.1 Instrument Defaults

The Multi-PRO 400 Instrument leaves the factory with the following settings:

Device Address

0

Parameters

1: DO concentration in mg/L
2: DO percent saturation

3: Temperature in

°C

4: Specific conductivity in µS/cm
5: Depth in feet
6: pH
7: ORP in millivolts

Output
Sequence

The default sequence is oTsCLHr.

o = Dissolved Oxygen Concentration
T = Temperature
s = Dissolved Oxygen Saturation
C = Specific Conductivity
L = Level
H = pH
r = ORP