2.4 Extended Commands

The Multi-PRO 400 Instrument supports the following extended SDI-12 commands.

Name

Command

Response

Set Factory Defaults




Send Data

aXFD!




aD0!

a0011<CR><LF>
One result is available in 1 second for reading by
the Send Data command. Restores all settings and
calibration values to their factory defaults.

a+s<CR><LF>
s = command status, 1 = command successful, 0 =
command failed.

Set Output Sequence

aXOnnn!

aOK<CR><LF> or
aERROR<CR><LF>

nnn = 1 to 15 parameter characters in the desired
output order. The default sequence is oTsCLHr.

o = Dissolved Oxygen Concentration
T = Temperature
s = Dissolved Oxygen Saturation
p = Partial Pressure of Oxygen

A = Actual Conductivity
C = Specific Conductivity
S = Salinity
D = Total Dissolved Solids
R = Resistivity
d = Density of Water

P = Pressure
L = Level

H = pH
h = pH mV
r = ORP

Read Output Sequence

aXO!

annn<CR><LF>
nnn = 1 to 15 parameter characters as described
for the XO command.

Set DO Concentration Units

aXoUnnn!

aOK<CR><LF> or
aERROR<CR><LF>

nnn = the DO concentration units ID
Three digits are required.

Read DO Concentration
Units

aXoU!

annn<CR><LF>
nnn = DO concentration units ID