3 calibration registers—ph/orp, Calibration registers—ph/orp – In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual

Page 52

Advertising
background image

Register

Size

(registers)

Mode & Access Level

(R/W)

Data

Type

Description

4557

1

R1/W2

ushort Units ID

162 = mV (default)

4558

1

R1

ushort Data Quality ID

4559

2

R1/W3

float

Off line sentinel value (default = 0.0)

4561

1

R1

16 bits Available Units = 0x0002 (2)

7.3 Calibration Registers—pH/ORP

These registers can be written only when the sensor is in the calibration mode. The probe will
return exception 0x92 (invalid sensor mode) if an attempt is made to write these registers when
the calibration mode is off. pH solution data must be written into the registers in order of
increasing pH.

Register

Size

(registers)

Mode & Access Level

(R/W)

Data

Type

Description

1618

1

R1/W3

ushort

Number of pH calibration points 1, 2, or

3 (default = 1)

1619

2

R1/W3

float

pH value of solution 1 (default = 0)

1621

2

R1/W3

float

mV response in solution 1 (default = 0)

1623

2

R1/W3

float

Temperature of solution 1, T

1

in °C

(default = 25)

1625

2

R1/W3

float

pH value of solution 2 (default = 0)

1627

2

R1/W3

float

mV response in solution 2, mV

2

(default = 0)

1629

2

R1/W3

float

Temperature of solution 2, T

2

in °C

(default = 25)

1631

2

R1/W3

float

pH value of solution 3 (default = 0)

1633

2

R1/W3

float

mV response in solution 3 (default = 0)

1635

2

R1/W3

float

Temperature of solution 3, T

3

in °C

(default = 25)

1637

2

R1/W3

float

Slope, m

1

in pH/mV

(default = -0.016903313)

1639

2

R1/W3

float

Offset, b

1

in pH (default = 7.0)

52