5.3 Calibration Registers—Conductivity

Values in the calibration registers determine how sensor parameters are calculated.

Register

Size

(registers)

Mode & Access Level

(R/W)

Data

Type

Description

0628

2

R1/W3

float

Cell Offset, K

0

(default = 0.0)

0630

2

R1/W3

float

Cell Constant, K

(default = 1.0)

0632

2

R1/W3

float

Reference Temperature,

T

ref

in °C (default = 25)

0634

2

R1/W3

float

Alpha Coefficient

α

(default = 0.0191)

0636

2

R1/W3

float

Beta Coefficient

β

0

(default = 1.0)

0638

2

R1/W3

float

Beta Coefficient

β

1

(default = 0.0)

0640

2

R1/W3

float

Beta Coefficient

β

2

(default = 0.0)

0642

2

R1/W3

float

Beta Coefficient

β

3

(default = 0.0)

0644

2

R1/W3

float

Beta Coefficient

β

4

(default = 0.0)

0646

2

R1/W3

float

Beta Coefficient

β

5

(default = 0.0)

0648

2

R1/W3

float

Beta Coefficient

β

6

(default = 0.0)

0650

2

R1/W3

float

Beta Coefficient

β

7

(default = 0.0)

0652

2

R1/W3

float

TDS Conversion Factor
CF

TDS

in ppm

(software limited range 0.001 to
10.000, default = 0.65, resolution
0.001)

5.3.1 Cell Offset and Cell Constant

These values are used to calibrate conductivity to user standards. These registers shall only be
able to be written when the sensor is in the calibration mode. The probe will return exception
0x92 (invalid sensor mode) if an attempt is made to write these registers when the calibration
mode is off.

Actual conductivity (AC) is calculated as follows.

AC = K

0

+ K * AC

f

Where AC

f

is the actual conductivity value computed using the factory calibrated cell constant.

For a single point calibration, K

0

is set to zero.