1 warm-up time, Warm-up time – In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual

Page 44

Advertising
background image

Register

Size

(registers)

Mode & Access Level

(R/W)

Data

Type

Description

1014

3

R1/W2

time

Next user calibration

(0 = none required)

1017

1

R1

ushort Warm-up time in milliseconds = 800

1018

1

R1

ushort Fast sample rate = 800 milliseconds

1019

1

R1

ushort Number of sensor parameters (N) = 3

1020

1

R1/W3

ushort Alarm/warning parameter number

(1 – N, default = 1)

1021

1

R1/W3

16 bits Alarm and warning enable bits

(default = 0)

Bit 0 = High alarm enabled

Bit 1 = High warning enabled

Bit 2 = Low warning enabled

Bit 3 = Low alarm enabled

Bit 4 = Sensor calibration warning

1022

2

R1/W3

float

High alarm set value (default = 0.0)

1024

2

R1/W3

float

High alarm clear value (default = 0.0)

1026

2

R1/W3

float

High warning set value (default = 0.0)

1028

2

R1/W3

float

High warning clear value

(default = 0.0)

1030

2

R1/W3

float

Low warning clear value

(default = 0.0)

1032

2

R1/W3

float

Low warning set value (default = 0.0)

1034

2

R1/W3

float

Low alarm clear value (default = 0.0)

1036

2

R1/W3

float

Low alarm set value (default = 0.0)

6.1.1 Warm-up Time

If automatic density correction is disabled, this register returns the level sensor warm-up time as
shown. If automatic density correction is enabled, this register returns the sum of the level
sensor and conductivity sensor warm-up times.

44