Advertising
background image

3. Modbus Registers—Probe

3.1 Modbus Communication Setup

See Connect the Communication Device on page 25. Click the Modbus setup button and assign
instrument settings according to the requirements of the controller.

The following Modbus registers are specific to the instrument. More information about Modbus,
including protocol specifications can be downloaded from

www.modbus.org

.

Register

Size

(registers)

Mode & Access Level

(R/W)

Data

Type

Description

7000

1

R1

ushort Probe Register Map Template Version (1)

7001

1

R1/W3

ushort Automatic Salinity Correction

7002

1

R1/W3

ushort Automatic Barometric Pressure Correction

7003

1

R1/W3

ushort Automatic Density Correction

7004

1

R1

ushort External Power Voltage (millivolts)

3.2 Automatic Salinity Correction

Set this register to 0 to disable automatic salinity correction. Set to 1 to enable automatic salinity
correction. The default is disabled. If disabled, the RDO sensor shall apply its live salinity
register value to correct its oxygen concentration for salinity. If enabled, the RDO sensor shall
use the salinity value from the conductivity sensor to correct its oxygen concentration for
salinity.