In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual

Page 18

Advertising
background image

Name

Command

Response

Set Temperature Units

aXTUnnn!

aOK<CR><LF> or
aERROR<CR><LF>

nnn = the temperature units ID
Three digits are required.

Read Temperature Units

aXTU!

annn<CR><LF>
nnn = temperature units ID

Restore DO Factory
Calibration




Send Data

aXoFC!





aD0!

a0011<CR><LF>
One result is available in 1 second for reading by
the Send Data command. Restores DO sensor
slope and offset calibration values to their factory
defaults. No change to other parameter settings.

a+s<CR><LF>
s = command status, 1 = command successful, 0 =
command failed.

Set DO Slope

aXoSpd.d!

aOK<CR><LF> or
aERROR<CR><LF>

pd.d = slope
p = polarity sign (+ or -)
d = digits (1 to 7)
. = decimal point (optional)

Read DO Slope

aXoS!

apd.d<CR><LF>
pd.d = slope

Set DO Offset

aXoOpd.d!

aOK<CR><LF> or
aERROR<CR><LF>

pd.d = offset
p = polarity sign (+ or -)
d = digits (1 to 7)
. = decimal point (optional)

Read DO Offset

aXoO!

apd.d<CR><LF>
pd.d = offset

Set Conductivity Units

aXCUnnn!

aOK<CR><LF> or
aERROR<CR><LF>

nnn = the actual and specific conductivity units ID
Three digits are required.

Read Conductivity Units

aXCU!

annn<CR><LF>
nnn = actual and specific conductivity units ID

18