1. Controller Requirements and Connections

The Multi-PRO 400 may be connected to a controller or logger for communication via:

• SDI-12
• RS485 Modbus
• RS232 Modbus (with converter)

1.1 Wiring Overview

Refer to the diagrams on the following pages. Trim and insulate unused wires. The shielded
wire should be wired to a chassis ground or earth ground.

Cable Stripped and Tinned

Signal

Color

Ground/Return

Black

External Power

Red

No Connection

Brown

RS485 (-)

Green

RS485 (+)

Blue

SDI-12

White

1.2 Power Connections

The Multi-PRO 400 requires an external 8 to 36 VDC power source. The red wire must be
connected to the positive terminal of the power source. The black wire must be connected to the
negative terminal of the power source, which is often referred to as the system ground or return.